Specifications

CRANEX D dimensions

 

CRANEX® D General Specifications
Generator High frequency DC generator, operating frequency 40 kHz
X-ray tube OPX 105 or equivalent
Focal spot size 0.5 mm
Minimum total filtration 2.7 mm Al
Line voltage 230/240 Vac ±10% / 115 Vac (50/60 HZ)
Anode voltage 57 - 85 kV
Anode current 10 mA
Exposure time 17.6 s panoramic, 8 - 20 s cephalometric
SID 520 mm (20.47”) panoramic, 1721 mm (67.75”) cephalometric
Fusing 8 A / 16 A slow (230/115 Vac)
Weight Pan 120 kg (264 lbs), Ceph 165 kg (363 lbs)
Electrical safety classification EN 60601-1 class 1/B
Color RAL 7040, RAL 9003



Digital Unit
Sensor CCD-detector
Active sensor surface PAN: 147.5 x 6.1 mm, CEPH: 221.2 x 6.1 mm
Pixel sizeμ 96 μm
Resolution 6,25 lp/mm
File size PAN max: 9.5 MB, CEPH max: 11.5 MB



Workstation computer recommendations
Operating system Windows® XP Professional/Home/SP1 or SP2, Windows® 2000 Professional/SP4